pH
● Color High contrast LCD touchscreen, 7.0 inches.
● Intelligent operation system provides features including user management, calibration management, method management, data management and etc.
● Multi-reading feature allows auto-read, timed-read and continuous-read.
● Automatic/Manual temperature compensation ensures accurate results.
● Data Storage 1000 sets for each parameter (GLP-compliant).
● Data analysis feature helps user review, compare and recalculate results.
● Support for USB or RS-232 communication.
● Reset feature automatically resumes all settings back to factory default options.
● IP54 waterproof.
| Model |
M500T-A
|
|
| Parameters | pH/EC/ISE/DO/Temp. (mV/ORP/pX/Resistivity/TDS/Sal.) | |
| pH | Range | -2.000 to 20.000pH |
| Resolution | 0.1, 0.01, 0.001pH | |
| Accuracy | ±0.002 pH | |
| Calibration Points | Up to 6 | |
| Standard Customization | Yes | |
| Calibration Reminder | Yes | |
| Standard Recognition | NIST, GB, DIN buffers | |
| Slope Limit | Yes | |
| mV/ORP | Range | -2000.00 to 2000.00 mV |
| Resolution | 0.1, 0.01 mV | |
| Accuracy |
reading whichever is greater
|
|
| EH ORP Mode | Yes | |
| ORP Calibration Points | 1 custom point (relative mV) | |
| pX | Range | -2.000 to 20.000 pX |
| Resolution | 0.1, 0.01, 0.001 pX | |
| Accuracy | ±0.002 pX | |
| Calibration Points | Up to 6 | |
| ISE | Range | 1.000e-9 to 9.999e+9 |
| Units |
mol/L, mmol/L, g/L, mg/L, μ g/L, ppm, ppb
|
|
| Resolution | Up to 4 significant digits | |
| Accuracy | ± 0.3% | |
| Calibration Points | Up to 6 | |
| Conductivity | Range | 0.000 μS/cm to 2000 mS/cm |
| Resolution | 0.001 μS/cm minimum, various with range selection | |
| Accuracy | ±0.5% FS | |
| Reference Temperature | 5, 10, 15, 18, 20, 25 ºC | |
| Calibration Points | Up to 5 | |
|
Calibration Reminder
|
Yes | |
| Standard Recognition |
10 μ S/cm, 84 μ S/cm, 500 μ S/cm, 1413 μ S/cm, 12.88 mS/cm; 146.5 μ S/
cm, 1408 μ S/cm, 12.85mS/cm, 111.3mS/cm
|
|
| Resistivity | Range | 5.00 Ω·cm~100.0 MΩ·cm |
| Resolution | 0.01 Ω·cm minimum | |
| Accuracy | ±0.5% FS | |
| TDS | Range |
0.000~1000 ppt
|
| Resolution | 0.001 ppm minimum, changed with range | |
| Accuracy | ±0.5% FS | |
| Salinity | Type | NaCl |
| Range |
0.0 ~ 80.0 ppt
|
|
| Resolution |
0.1ppt
|
|
| Accuracy |
± 1ppt
|
|
| Temperature | Range |
-10 to 135 ºC, 14 to 275 ºF
|
| Unit |
ºC, ºF
|
|
| Resolution | 0.1 | |
| Accuracy | ±0.1 | |
| Measurement | Reading Mode | Auto Read (Fast, Medium, Slow), Timed, Continuous |
| Reading Prompts | Reading, Stable, Locked | |
| Temp. Compensation | ATC, MTC | |
| Data management | Data Storage | 1000 Groups |
| GLP Features | Yes | |
| Log Management | Yes | |
| Inputs | pH Electrode | BNC(Q9) |
| DO with Temp. Probe |
4-pin aviation connector
|
|
| Conductivity with Temp. Probe |
5-pin aviation connector
|
|
| Outputs | USB | USB 2.0 flash memory device, PC,scanner, |
| RS 232 | printer,auto-sampler | |
| Display options | Backlight | Yes |
| Auto Shutdown | 1~60 min, off | |
| IP Rating | IP54 | |
| Date and Time | Yes | |
| General | Power | AC Adapter, 100-240V AC input, DC24V output |
| Dimensions | 220×195×68mm | |
| Weight | 950g (2.09 lb) | |