M600L
Product Price : $1552
pH
● 1-6 points calibration with Standard Recognition.
● Selectable pH buffer groups, including NIST, DIN, GB.
● Support for pH standard group customization.
● Automatic electrode diagnosis with pH slope and offset display.
Ion
● 1-6 points calibration.
● Selectable measurement unit, including μg/L, mg/L, g/L, mmol/L,
pX, etc.
● Multi-measurement modes are supported, including Direct Reading mode, Standard Addition mode, Sample Addition mode
and GRAN mode.
Conductivity
● 1-5 points calibration with Standard Recognition.
● Settable parameters, including cell constant, reference
temperature (5/10/15/18/20/25°C), temperature compensation
coefficient and TDS factor.
● Temperature compensation type (none, linear, pure water).
DO
● Support for air-saturated water or zero oxygen calibration.
● Auto barometric pressure compensation.
● Manual Salinity Factor Correction.
● Selectable pressure unit, including kPa, mbar, Torr, Atm.
● Color High contrast LCD touchscreen, 7.0 inches.
● Intelligent operation system provides features including usermanagement, calibration management, method management, data management and etc.
● Multi-reading feature allows auto-read, timed-read and continuous-read.
● Automatic/Manual temperature compensation ensures accurate results.
● Data Storage 1000 sets for each parameter (GLP-compliant).
● Data analysis feature helps user review, compare and recalculate results.
● Support for USB or RS-232 communication.
● Reset feature automatically resumes all settings back to factory default options.
● IP54 waterproof.
● Support for three modules, namely pH/ISE module, conductivity module and dissolved oxygen module.
● Select from three modules to customize the four channels your way.
● Four channels can be configured in any combination.
| Model | M600T | |
| Parameters | pH/EC/ISE/DO/Temp. (mV/ORP/pX/Resistivity/TDS/Sal./DO Saturation) | |
| pH | range | -2.000 to 20.000pH |
| Resolution | 0.1, 0.01, 0.001pH | |
| Accuracy | ±0.002 pH | |
| Calibration Points | Up to 6 | |
| Standard Customization | Yes | |
| Calibration Reminder | Yes | |
| Standard Recognition | NIST, GB, DIN buffers | |
| Slope Limit | Yes | |
| mV/ORP | Range | -2000.00 to 2000.00 mV |
| Resolution | 0.1, 0.01 mV | |
| Accuracy | ± 0.03% | |
| EH ORP Mode | Yes | |
| ORP Calibration Points | 1~2 custom point (relative mV) | |
| pX | Range | -2.000 to 20.000 pX |
| Resolution | 0.1, 0.01, 0.001 pX | |
| Accuracy | ±0.002 pX | |
| Calibration Points | Up to 6 | |
| ISE | Range | 1.000e-9 to 9.999e |
| Units | mol/L, mmol/L, g/L, mg/L, μ g | |
| Resolution | Up to 4 significant digits | |
| Accuracy | ± 0.3% | |
| Calibration Points | Up to 6 | |
| Conductivity | Range | 0.000 μS/cm to 3000 mS/cm |
| Resolution | 0.001 μS/cm minimum, various with range selection | |
| Accuracy | ±0.5% FS | |
| Reference Temperature | 5, 10, 15, 18, 20, 25 ºC | |
| Calibration Points | Up to 5 | |
| Standard Recognition | 146.5 μ S/cm, 1408 μ S/cm, 12.85mS/cm, 111.3mS/cm | |
| Resistivity | Range | 5.00 Ω·cm~100.0 MΩ·cm |
| Resolution | 0.01 Ω·cm minimum | |
| Accuracy | ±0.5% FS | |
| TDS | Range | 0.000 mg/L~1000 g/L |
| Resolution | 0.001 ppm minimum, changed with range | |
| Accuracy | ±0.5% FS | |
| Salinity | Type | NaCl% |
| Range | (0.0 ~ 8.00 )% | |
| Resolution | 0.01% | |
| Accuracy | ± 0.1% | |
| Dissolved | Range | 0.00 to 99.99 mg/L |
| Oxygen Concentration - Polarographic | Resolution | 0.01 mg/L |
| Accuracy | ± 0.1 mg/L | |
| Calibration Reminder | Yes | |
| Calibration Points | Air-saturated water or zero point | |
| Barometric Compensation | Yes | |
| Manual Salinity Factor Correction | Yes | |
| % Saturation -Polaro- graphic | Range | (0.0 to 600)% |
| Resolution | 0.10% | |
| Accuracy | ±2.0% | |
| Temperature | Range | -5 to 130 ºC |
| Unit | ºC | |
| Resolution | 0.1 | |
| Accuracy | ±0.1 | |
| Measurement | Reading Mode | Auto Read (Fast, Medium, Slow), Timed, Continuous |
| Reading Prompts | Reading, Stable, Locked | |
| Temp. Compensation | ATC, MTC | |
| Data management | Data Storage | 1000 Groups |
| GLP Features | Yes | |
| Log Management | Yes | |
| Inputs | pH Electrode | BNC(Q9) |
| DO with Temp. Probe | 6-pin Mini-DIN | |
| Conductivity with Temp. Probe | 8-pin Mini-DIN | |
| Outputs | USB,RS 232 | USB 2.0 flash memory device, printer,PC, |
| Display options | Backlight | Yes |
| Auto Shutdown | 1~60 min, off | |
| IP Rating | IP54 | |
| Date and Time | Yes | |
| General | Power | AC Adapter, 100-240V AC input, DC24V output |
| Dimensions | 280×280×130mm | |
| Weight | 2500g (5.51 lb) | |
We use cookies to collect information about how you use this site. We use this information to make the website work as well as possible and improve our services.more details