The SX-Z Hamick drum is used to measure the amount of particulate matter emitted by clean fabrics during simulated motion. This motion involves the fabric in a dry state being flipped inside a rotating drum. Within a specified time, a certain amount of samples is taken to measure the quantity
or concentration of particulate matter emitted by the fabric. The measured value is then compared with the standard grade to determine the fabric's cleanliness level.
The principle of the test bench is based on the American Society for Environmental Science and Technology standard "IEST-RP-CC003.5" and "Clothing Elements for Clean Rooms and Other Controlled Environments", using the Hammett drum mode.
Function
- Sampling flow rate: 28.3L/min ±5% (1 CFM)
- Particle size distribution error: ±30%FS
- Error of indicated concentration: ±30%FS
- Self-cleaning time: less than 10min
- Relative error of repeatability: less than or equal to 10%FS
- Complies with GMP static, GMP dynamic and ISO14644-1 standards
- Sampling concentration: 35,000 particles/l
- 5000 stored data
- Roller speed: 1~20rpm adjustable
Parameters